Electric Holding Furnace Chitherm Hrf512-05n Nitrogen Dryer Consistent Ceramic Drying Solution
1. Applications:
Mainly used for binder burning out(BBO) process of electronic components, ceramic substrates.
2. Mainly Specifics:
- Rated Temp.: 350ºC
- Max Temp.: 500ºC
- Effective Dimensions: 800 × 800 × 800mm(W×H×D)
- Heating Method: Stainless Steel Heater
- Process Atmosphere: Nitrogen
- Temperature Control Stability: ±1ºC, with PID parameter self-tuning function
- Temperature Uniformity: better than ±5ºC (insulation 1 hour at 400ºC test)
- Thermocouple: Type K
- Temperature control points: 1
- Temperature measurement points: 1 point
- Process Steps: 4 × 10 steps
- Exhaust: One exhaust port on top
- Alarm System: over temperature, thermocouple failure and other abnormal conditions will activate sound and light alarm protection
- Max Heating Power: 24kW
- Insulation Power: ≤ 12kW
- Surface Temperature Rise: ≤ 35ºC
- Weight: about 1000kg
- Furnace Dimensions: about 1500×1930×2000mm (W×H×D);
- Appearance Color: light gray
3. Delivery List |
|
Item |
Note |
QTY |
Basic Compositions |
Dryer |
|
1 PC |
Certificates of Inspection |
Dryer and major purchased components |
1 Set |
Technical Documents |
Manual instructions, technical documents of major purchased components, etc |
1 Set |
Key Parts |
Heating Elements |
Stainless Steel Heater |
1 Set |
Thermocouple |
WATLOW/THERMOWAY, Type K |
1 PC |
Temperature Controller |
|
1 Set |
Monitor |
Touch screen |
1 Set |
Spares |
SSR |
|
1 PC |
4. Facility Requirements
4.1 Environmental conditions: Temperature 0 - 40ºC, humidity ≤ 80% RH, no corrosive gas, no strong
airflow disturbance.
4.2 Process air conditions: 99.999% purity nitrogen, pressure 0.2 - 0.4Mpa, gas consumption 10m3 /h.
4.3 Ventilation system: non-contact access to the user pumping system, pumping capacity is greater
than 15m3 /h;
4.4 Ground requirements: level, no obvious vibration, bearing capacity >300Kg/m2 .
4.5 Power conditions: capacity greater than 32kVA, 3 phases 5 lines, voltage 220/380V, frequency
50Hz(according to local situation). Live wire: yellow, green, red, neutral wire: blue, ground wire: yellow-green;
4.6 Installation site: 2000mm×3000mm×3000mm (W×H×D), installation area greater than 6m2.